• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE ANGIOJET®; CATHETER, EMBOLECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE ANGIOJET®; CATHETER, EMBOLECTOMY Back to Search Results
Model Number UNK872
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/08/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: it is indicated that the device will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.The batch number is unknown and the manufacturing records for the complaint device could not be reviewed.The most probable root cause was unable to be determined.(b)(4).
 
Event Description
Same case as: 2134265-2017-02514.It was reported that the catheter ruptured during use.Three unspecified angiojet® catheters and an angiojet® ultra system console were selected for a thrombectomy procedure in the iliac vein.The first two catheters were unable to prime and the pump filled with water and broke.A third catheter was able to prime; however after 49 seconds in thrombectomy mode, it ruptured.The saline bags and cables were changed but the issue persisted.The case did not continue and the patient stayed overnight.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET®
Type of Device
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6427042
MDR Text Key70633853
Report Number2134265-2017-02515
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNK872
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/08/2017
Initial Date FDA Received03/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-