PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 INLAY OPTIMA MULTI-LENGTH URETERAL STENT WITHOUT GUIDEWIRE, 6 FR. Back to Search Results
Catalog Number 788600
Device Problem Knotted (1340)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that when the stent was removed from the patient, it was tied in a knot.There was no patient injury and no medical intervention required.There was no delay to the procedure reported.The surgeon stated that the stent was knotted up in the bladder upon removal.
 
Manufacturer Narrative
Received 1 used stent only.The reported event was confirmed; however, the cause is unknown.During the visual inspection noted that stent appeared to be tied into a knot.Per dimensional evaluation: length = 8.718 in (specification is 8.66 in ± 0.39 in).The sample was found within specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: " contraindications: no known contraindications for use.Precautions: suture may be cut off prior to stent placement.Remove suture if indwelling time is expected to be longer than 14 days.Avoid improper handling of stent such as bending, kinking, tearing, etc.Misuse could damage the overall integrity of the stent.Ureteral stents should be checked periodically for signs of encrustation and proper function.Periodic checks of the stent by cystoscopic and/or radiographic procedures are recommended at intervals deemed to be appropriate by the physician in consideration of the individual patient¿s condition and other patient specific factors.When long-term use is indicated, it is recommended that indwelling time not exceed 365 days.The stent is not intended as a permanent indwelling device.With any ureteral stent, migration is a possible complication, which could require medical intervention for removal.Selection of too short a stent may result in migration.Care should be exercised when removing the stent from the inner polybag to eliminate tearing or fragmentation.The insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure.Directions for use: determine the proper stent length for the patient.This is generally calculated from the baseline pyelogram.Accurate measurements will optimize drainage efficiency and patient comfort.Insert the cystoscope then pass the guidewire* through the scope until the tip is in the renal pelvis.Move the pigtail straightener over the proximal end (kidney coil end) of the ureteral stent allowing easier insertion onto the guidewire.Remove pigtail straightener once the stent is secure on the guidewire.Pass the stent over the guidewire through the cystoscope by using the push catheter for proper placement.Watch the distal end (bladder coil end) of the stent or the radiopaque, proximal end of the pusher.Stop advancing when the stent¿s distal end marker reaches the ureterovesical junction (uvj).(see below for proper placement directions on the multi-length ureteral stent.) withdraw the guidewire slowly.The stent will form a pigtail automatically.Carefully remove the push catheter.Activate the guidewire coating according to the ¿instructions for use¿ found within the guidewire packaging.Multi-length ureteral stent placement: to accurately size this stent count the marker bands as it is being advanced into the ureter.The first large band indicates the 22cm length.The second and third bands indicate 24cm and 26cm lengths respectively.The last large band is the 28cm length.If you need to place for the 30cm and 32cm lengths, use the attached suture or endoscopic forceps to gently pull back on the stent unwinding the coil from the kidney.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and with applicable local, state and federal laws and regulations.This is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient." (b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that upon removal of the stent, it was noted that the device was in a knot.There was no patient injury and no medical intervention required.There was no delay to the procedure reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INLAY OPTIMA MULTI-LENGTH URETERAL STENT WITHOUT GUIDEWIRE, 6 FR.
Type of Device
URETERAL STENT
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX   85621
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6429472
MDR Text Key70981522
Report Number1018233-2017-01296
Device Sequence Number1
Product Code FAD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043193
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/16/2020
Device Catalogue Number788600
Device Lot NumberNGAV2743
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/02/2017
Initial Date FDA Received03/23/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-