After the reported event, the affected device was checked by a draeger service technician who also downloaded the electronic log file.Based on the information logged therein the case in question was started on (b)(6), at 2:42p.M.Until 3:10p.M., ventilation was stable in both volume- and pressure controlled modes with a constant peep of 5mbar.From 3:10p.M., the peep value increased to 8mbar.Although this may have been the result of user settings, this is rather the timeframe when the reported problems occurred.Within his device check, our service technician was able to reproduce the reported slow pressure decrease during expiration and determined that it was due to a sporadically sticking peep valve.During further inspection of the breathing system neither moisture nor any other contamination could be found, which could have restricted the function of the peep valve.The problem could be solved by re-assembling the peep valve assembly according to the latest draeger instructions which ensure that the diaphragm of the peep valve is mounted free of tension.Thus it was concluded that the reported event was due to unwanted tension within the peep valve diaphragm, which can result from an unfavorable assembly of the peep valve.In this context draeger introduced improved assembly instructions end of 2012.The affected device has already been repaired accordingly.The amount of similar cases is significantly below the values that provide basis for our risk management.Consequently, no further actions are required at this time.
|