MEDTRONIC NEUROMODULATION ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problems Failure to Interrogate (1332); Communication or Transmission Problem (2896); Human-Device Interface Problem (2949)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/23/2017
Event Type  malfunction  
Event Description
A manufacturing representative (rep) reported that a surgeon in the operating room (or) was changing out a device and they were unable to communicate with the new device.The clinician programmer would connect with the old device but not with the new one.They tried communicating while the device was in the pocket and out of the pocket.They tried turning the surgical lights away from the device to avoid interference.The rep stated they would follow up with the hcp.No further complications are anticipated.
 
Event Description
Additional information from the rep reported they were able to resolve the issue by deleting previous programming session entrees in the programmer, thus freeing up memory.The surgery proceeded without further incident and all was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer Contact
lisa clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6436230
MDR Text Key70988552
Report Number3007566237-2017-01137
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 03/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/23/2017
Initial Date FDA Received03/27/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-