Model Number CENTRALINK DATA MANAGEMENT SYSTEM
Device Problem Device Issue (2379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/28/2017
Event Type  malfunction  
Manufacturer Narrative
A siemens headquarters support center (hsc) specialist reviewed the event.Hsc found the cerberus connection timed out before the upgrade.This caused the images to not transfer immediately.The centralink service was rebooted ten (10) times.After the last reboot, the server started and an image was accepted.The cause of the delay in receiving hematology images was due to the cerberus connection timed out before the upgrade.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Event Description
The customer reported that images from their hematology instruments, from a remote site, were delayed after an update and reboot of the centralink system.The application cerberus, did not see the hospital connection, which is what is used by the remote site to send files.There are no known reports of patient intervention or adverse health consequences due to the delay in receiving hematology images.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRALINK DATA MANAGEMENT SYSTEM
Type of Device
DATA MANAGER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict ave.
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICSMANUFACTURING LTD
registration #: 8020888
chapel lane
swords, co, dublin 10040
EI   10040
Manufacturer Contact
timothy rice
511 benedict ave
tarrytown, NY 10591
9145242406
MDR Report Key6437860
MDR Text Key70990916
Report Number2432235-2017-00221
Device Sequence Number1
Product Code JQP
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCENTRALINK DATA MANAGEMENT SYSTEM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/02/2017
Initial Date FDA Received03/27/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-