• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. H/S CUVETTE 1/2X1/2-JAPAN; BLOOD GAS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. H/S CUVETTE 1/2X1/2-JAPAN; BLOOD GAS MONITOR Back to Search Results
Model Number CV-6912
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 11/14/2016
Event Type  malfunction  
Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup, an error message "h/s disconnect at cuvette" appeared when the cdi h/s cuvette was connected to the cdi h/s probe.The measurement was also not available.This event is associated with a voluntary safety alert submitted by terumo on december 8, 2015.The safety alert number is 1124841-12/08/2015-004-c.No patient involvement as this occurred during set up.Product was changed out.The procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to the fda in accord with applicable regulations - and as indicated by terumo cardiovascular system in the initial report submitted to the fda on march 28, 2017.(b)(4).The actual sample was visually inspected, during which no anomalies were noted.A review of the device history record revealed no manufacturing anomalies.The sample was microscopically inspected, and the magnet of the cuvette did not reveal any potential defect that would inhibit part functionality.The sample was found to have no difficulties connecting to the cdi 500, and the magnet strength was found to be within specifications.A definitive root cause could not be determined; therefore, this event is not confirmed.This event could be associated with a voluntary safety alert submitted by terumo on december 8, 2015.The safety alert number is 1124841-12/08/2015-004-c.All available information has been placed on file in quality management for appropriate tracking, trending and follow up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
H/S CUVETTE 1/2X1/2-JAPAN
Type of Device
BLOOD GAS MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
125 blue ball road
elkton MD 21921
Manufacturer Contact
cathleen hargreaves
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key6438786
MDR Text Key71053947
Report Number1124841-2017-00047
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915265
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberCV-6912
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Other Device ID Number(01)00699753270114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/07/2017
Initial Date FDA Received03/28/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/12/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1124841-12/08/2015-004-C
Patient Sequence Number1
-
-