• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION ZM-541PA; TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION ZM-541PA; TRANSMITTER Back to Search Results
Model Number ZM-541PA
Device Problems Melted (1385); Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/02/2017
Event Type  malfunction  
Manufacturer Narrative
The customer reports that the transmitter got hot and melted the battery compartment.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Further evaluation being performed.
 
Event Description
The customer reports that the transmitter got hot and melted the battery compartment.
 
Manufacturer Narrative
Manufacturer narrative: the customer reports that the transmitter got hot and melted the battery compartment.The unit was cleaned and evaluated.The reported problem of getting hot was not duplicated.However, it was found that the spo2 connector was broken and the non-invasive blood pressure (nibp) function failed.All malfunctioning parts have been replaced.The unit was tested per the operator's/service manual and the results were recorded on the maintenance check sheet.The unit completed 24 hours of extended testing and operates to manufacturer's specifications.Nihon kohden will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-541PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, japan 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
gunma 161-8 560
JA   161-8560
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgm
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
2687488
MDR Report Key6443030
MDR Text Key71468835
Report Number8030229-2017-00087
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921106617
UDI-Public04931921106617
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043517
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/29/2017,03/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZM-541PA
Device Catalogue NumberZM-541PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/29/2017
Distributor Facility Aware Date03/02/2017
Device Age41 MO
Event Location Hospital
Date Report to Manufacturer03/29/2017
Initial Date Manufacturer Received 03/29/2017
Initial Date FDA Received03/29/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/18/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-