Model Number N/A
Device Problems Break (1069); Leak/Splash (1354)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 03/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reported device is not available for evaluation.The event is currently under investigation.
 
Event Description
An international customer reported that the stopcocks are breaking and / or leaking; it was stated "the neck has a tear." the event reportedly occurred prior to patient contact.No further information was provided.
 
Manufacturer Narrative
A review of the complaint history, drawings, device history record, manufacturing instructions, trends, quality control, and visual inspection of the returned device was conducted during the investigation.Thirty three-way plastic stopcocks were returned for investigation, twenty-nine unused and one used.Visual inspection revealed that the returned devices were cracked at the base of the stopcocks and leaked.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that would contribute to this failure mode.There were no other reported complaints for this lot number.A definitive root cause could not be determined.Measures have been conducted to address this failure mode.Monitoring will continue to be performed for similar complaints.Based on the information provided and the results of our investigation; a definitive root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREE-WAY PLASTIC STOPCOCK
Type of Device
KGZ ACCESSORIES, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
rita harden
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key6447041
MDR Text Key71336506
Report Number1820334-2017-00501
Device Sequence Number1
Product Code KGZ
UDI-Device Identifier00827002002197
UDI-Public(01)00827002002197(17)180101(10)4009999
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberPTWS-2FLL-MLL-R
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/02/2017
Initial Date FDA Received03/30/2017
Supplement Dates Manufacturer Received09/14/2017
Supplement Dates FDA Received10/17/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-