• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRAEGERWERK AG AND CO KGAA APOLLO ANESTHESIA MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRAEGERWERK AG AND CO KGAA APOLLO ANESTHESIA MACHINE Back to Search Results
Model Number APOLLO
Device Problems No Audible Alarm (1019); Device Operates Differently Than Expected (2913)
Patient Problems Low Blood Pressure/ Hypotension (1914); Low Oxygen Saturation (2477)
Event Date 02/16/2017
Event Type  No Answer Provided  
Event Description
Following open heart procedure, while pt still in the operating room and connected to the anesthesia machine, patient's sao2 and bp were dropping.At this time the physician and the staff did not hear any audible alarms being triggered on the anesthesia machine.No adverse effect to the patient as a result of this event.Equipment tested by (b)(6) hospital clinical tech, department as well as field service engineer of the mfr.Internal and external investigation deemed equipment to be functioning per mfr specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APOLLO ANESTHESIA MACHINE
Type of Device
APOLLO ANESTHESIA MACHINE
Manufacturer (Section D)
DRAEGERWERK AG AND CO KGAA
MDR Report Key6453138
MDR Text Key71689440
Report NumberMW5068848
Device Sequence Number1
Product Code BSZ
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 03/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberAPOLLO
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/30/2017
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age74 YR
Patient Weight100
-
-