MEDTECH SA ROSA SURGICAL DEVICE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA 2.5.8
Device Problem Data Problem (3196)
Patient Problem No Code Available (3191)
Event Date 08/05/2015
Event Type  Injury  
Manufacturer Narrative
The device ro 14 043 has been inspected for investigation purpose.The tests performed confirmed that the registration has been lost when the user decided to overwrite the patient folder saved on the system with the one imported from the usb stick.The internal complaint reference is the following: (b)(4).
 
Event Description
It has been reported that during a surgery, a software failure has been detected.The robot experienced a collision between the instrument adapter and anchor bolt and the surgeon was forced to shutdown the device.Upon reboot, the registration and plan were no longer available in the patient folder.A surgery delay has been reported of 2 hours.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA SURGICAL DEVICE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
7414400
MDR Report Key6453208
MDR Text Key71495085
Report Number3009185973-2017-00080
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Inspection
Type of Report Initial
Report Date 03/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberROSA 2.5.8
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/05/2015
Initial Date FDA Received04/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-