The reporter indicated that the surgeon implanted a 13.2mm vicmo13.2 implantable collamer lens, -9.0 diopter, in the patient's right eye (od) on (b)(6) 2017.The patient experienced headache, elevated iop.The customer states there was no center hole on the lens.On (b)(6) 2017 the lens was exchanged and the problem was resolved.As of the day of mdr submission, the lens has not been returned.
|
Device evaluation: the lens was returned in small vial, there was moisture on the lens.Visual inspection found no visible damage, no hole on the lens, clear surgical residue on lens surface, and there was ink staining on the surface of the lens.The lens was remeasured, and passed the lens verification.(b)(4).
|