Lot Number 4627659
Device Problems Component Falling (1105); No Apparent Adverse Event (3189)
Patient Problems No Consequences Or Impact To Patient (2199); Foreign Body In Patient (2687); No Code Available (3191)
Event Date 03/29/2017
Event Type  Injury  
Manufacturer Narrative
Pending device return and engineering evaluation.
 
Event Description
After torquing the screw, when pulling the driver out, the head of the screw fell into the wound site and had to be retrieved.
 
Manufacturer Narrative
It is noted that only qualified surgeons who are fully knowledgeable about all aspects of surgical technique and are fully trained to use the system and instrumentation are to use these devices.The most likely cause of the driver failing to retain the screw appears to be due to the surgeon not fully pushing the driver completely over the poly plugs.This device is used for treatment not diagnosis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORQUE DEFINING SCREW KIT
Type of Device
SCREW KIT
Manufacturer (Section D)
EXACTECH, INC
2320 nw 66th ct
gainesville FL 32653
MDR Report Key6456225
MDR Text Key71598986
Report Number1038671-2017-00213
Device Sequence Number1
Product Code KWS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 12/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date10/17/2021
Device Lot Number4627659
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/30/2017
Initial Date FDA Received04/04/2017
Supplement Dates Manufacturer Received12/04/2018
Supplement Dates FDA Received12/18/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age70 YR
-
-