Model Number M0068318261 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/13/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a capio slim was used during an anterior repair procedure on (b)(6) 2017.According to the complainant, the needle detached from the suture.The needle was found inside the capio cage.The procedure was completed with another capio suturing device and suture.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
|
Manufacturer Narrative
|
A visual examination of the returned capio slim revealed that the cage was damaged.Functional inspection found that the carrier extends and retracts without any problem.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.
|
|
Event Description
|
It was reported to boston scientific corporation that a capio slim was used during an anterior repair procedure on (b)(6) 2017.According to the complainant, the needle detached from the suture.The needle was found inside the capio cage.The procedure was completed with another capio suturing device and suture.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
|