• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MARLBOROUGH SYMPHION¿ TISSUE REMOVAL SYSTEM FLUID MANAGEMENT ACCESSORIES; INSUFFLATOR, HYSTEROSCOPIC, FLUID, CLOSED-LOOP RECIRCULATION WITH CUTTER-COAGULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MARLBOROUGH SYMPHION¿ TISSUE REMOVAL SYSTEM FLUID MANAGEMENT ACCESSORIES; INSUFFLATOR, HYSTEROSCOPIC, FLUID, CLOSED-LOOP RECIRCULATION WITH CUTTER-COAGULA Back to Search Results
Model Number FG-0202
Device Problem Component or Accessory Incompatibility (2897)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/24/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a symphion fluid management accessories was used during a polypectomy procedure performed on (b)(6) 2017.According to the complainant, during preparation, the fluid management accessories was being set-up and the pressure sensor was not detected.The scope was disconnected from the controller several times and the controller was restarted twice and the pressure sensor was still not detected.A new fluid management accessory was used with the same scope and controller and the procedure was completed.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be fine.
 
Manufacturer Narrative
A symhpion device was returned for analysis.Visual analysis revealed no signs of damage to the device.Functional analysis revealed that the pressure sensor connector was possible to be inserted into the receptacle on the connector.However, the controller did not detect the pressure sensor.Therefore, the complaint was confirmed.Given the event description and condition of the returned device, there isn't enough information to determine a probable root cause.A review of the device history record (dhr) was performed; no anomalies were noted.A search of the complaint database revealed that no similar complaints exist for the specified lot.
 
Event Description
It was reported to boston scientific corporation that a symphion fluid management accessories was used during a polypectomy procedure performed on (b(6) 2017.According to the complainant, during preparation, the fluid management accessories was being set-up and the pressure sensor was not detected.The scope was disconnected from the controller several times and the controller was restarted twice and the pressure sensor was still not detected.A new fluid management accessory was used with the same scope and controller and the procedure was completed.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMPHION¿ TISSUE REMOVAL SYSTEM FLUID MANAGEMENT ACCESSORIES
Type of Device
INSUFFLATOR, HYSTEROSCOPIC, FLUID, CLOSED-LOOP RECIRCULATION WITH CUTTER-COAGULA
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6460740
MDR Text Key72097701
Report Number3005099803-2017-01075
Device Sequence Number1
Product Code PGT
UDI-Device Identifier08714729863717
UDI-Public(01)08714729863717(11)20180102(10)0058639701
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141848
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/02/2018
Device Model NumberFG-0202
Device Lot Number0058639701
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/24/2017
Initial Date FDA Received04/05/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/17/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age39 YR
-
-