|
Catalog Number RBY2C0740 |
Device Problems
Bent (1059); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/09/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Results: the pusher assembly and introducer sheath of the first ruby coil evaluated were kinked in multiple locations.Conclusions: evaluation of the returned devices revealed the first ruby coil had multiple kinks in the pusher assembly and introducer sheath.This damage was likely incidental and likely occurred during packaging for return.Due to this incidental damage, the locations and root cause of the reported kink could not be determined.Further evaluation revealed the second ruby coil was not detached.Therefore, the root cause of the reported unintentional detachment could not be determined.The microcatheter mentioned in the complaint was not returned for evaluation.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Event Description
|
The patient was undergoing a coil embolization procedure using ruby coils.During the procedure, prior to use, the hospital staff inadvertently bent a ruby coil pusher assembly while removing the ruby coil from its dispenser hoop.Consequently, the ruby coil would not advance into the microcatheter and therefore, it was set aside.While attempting to load a new ruby coil into the microcatheter, the hospital staff noticed that the ruby coil had unintentionally detached from its pusher assembly and therefore, the ruby coil was set aside.The procedure was completed using additional ruby coils and the same microcatheter.There was no report of an adverse effect to the patient.
|
|
Search Alerts/Recalls
|
|
|