• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - DOMINICAN REPUBLIC CLEARLINK BLOOD RECIPIENT SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - DOMINICAN REPUBLIC CLEARLINK BLOOD RECIPIENT SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8750
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 03/13/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an access set failed to engage the luer locking mechanism.The product was being flushed with saline solution at the time.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was returned and an evaluation is complete.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Functional test was performed and the device failed to engage the luer locking mechanism.The reported condition was verified; the cause is unknown.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK BLOOD RECIPIENT SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina san cristobal
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina san cristobal
DR  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6465157
MDR Text Key72127697
Report Number1416980-2017-03071
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/06/2021
Device Catalogue Number2C8750
Device Lot NumberDR16L06062
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/14/2017
Initial Date FDA Received04/06/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/04/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N