SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+ USA Back to Search Results
Catalog Number 10379675
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem Not Applicable (3189)
Event Date 03/09/2017
Event Type  malfunction  
Manufacturer Narrative
The customer returned the instrument for investigation as well as the power supply.Investigation of the instrument determined that there was a failure of d16 on the main pcb.D16 is the overvoltage protector diode, which blows if the voltage applied to instrument is too high.The power supply was tested and determined to be performing as expected.The root cause of the failure is not able to be determined.
 
Event Description
The customer reported smoke coming from the back of the clinitek status+.There was no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+ USA
Type of Device
CLINITEK STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key6465248
MDR Text Key72338408
Report Number3002637618-2017-00064
Device Sequence Number1
Product Code JIL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10379675
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/09/2017
Initial Date FDA Received04/06/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-