MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRI-SIL I. C. 2 WAY ALL-SILICONE FOLEY CATHETER WITH 5CC BALLOON.; FOLEY CATHETER (SILICONE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRI-SIL I. C. 2 WAY ALL-SILICONE FOLEY CATHETER WITH 5CC BALLOON.; FOLEY CATHETER (SILICONE) Back to Search Results
Catalog Number 1758SI14
Device Problems Fluid/Blood Leak (1250); Mushroomed (2987)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/30/2016
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that during the pretest, the catheter leaked water.The balloon was allegedly damaged, and a mushroom/ridge was identified.It was later reported, that an in-house observation, noted no damage to the balloon of the returned sample.There were no missing pieces.
 
Manufacturer Narrative
Received 1 silicone catheter.The reported event was confirmed as cause unknown based on the preliminary visual evaluation.Per the visual evaluation of the returned sample, a cuff roll was not observed.During the functional evaluation, a cuff roll was not formed.The dimensional evaluation of the catheter was found within specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "caution: do not aspirate urine through drainage funnel wall.Note: aggressive traction, particularly in the presence of suturing, is not recommended for 100% silicone foley catheters.Recommended inflation capacities: 3cc balloon: use 5cc sterile water.5cc balloon: use 10cc sterile water.30cc balloon:use 35cc sterile water.Do not exceed recommended capacities.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Visually inspect the product for any imperfections or surface deterioration prior to use." (b)(4).
 
Event Description
It was reported that during the pretest, the catheter leaked water.The balloon was allegedly damaged, and a mushroom/ridge was identified.It was later reported, that an in-house observation, noted no damage to the balloon of the returned sample.There were no missing pieces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX LUBRI-SIL I. C. 2 WAY ALL-SILICONE FOLEY CATHETER WITH 5CC BALLOON.
Type of Device
FOLEY CATHETER (SILICONE)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
lot 57c
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
lot 57c
covington GA 30014
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6466071
MDR Text Key72252977
Report Number1018233-2017-01584
Device Sequence Number1
Product Code MJC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K984136
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/12/2020
Device Catalogue Number1758SI14
Device Lot NumberNGZF3481
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/02/2016
Initial Date FDA Received04/06/2017
Supplement Dates Manufacturer Received12/08/2017
Supplement Dates FDA Received12/12/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/14/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-