EDWARDS LIFESCIENCES VFEM FEMORAL VENOUS CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number VFEM022 |
Device Problem
Break (1069)
|
Patient Problems
Death (1802); Blood Loss (2597)
|
Event Date 03/16/2017 |
Event Type
Death
|
Manufacturer Narrative
|
Additional manufacturer narrative: the device was not returned to edwards for evaluation.The clinical observation was unable to be confirmed.Manufacturing records were not reviewed as the lot number was not provided.Base on the information received, a definitive root cause could not be determined.The instructions for use (ifu) was reviewed and no inadequacies were identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The ifu provides the following: "contraindications for use femoral access cannulae are contraindicated for long-term use (> 6 hrs), including extra corporeal membrane oxygenation (ecmo) procedures." no further corrective or preventative actions are required at this time.Trends will continue to be monitored through the use of edwards quality systems and if action is required, appropriate investigation will be performed.
|
|
Event Description
|
Edwards received information that the subject cannula was inserted into this patient for ecmo treatment for 21 days.The patient was on the waiting list for lung transplantation.On the 21st day, the femoral cannula suddenly separated at the overmold section.Blood loss was noticed at the separation area which lead to the deterioration of the patient¿s condition.Then the patient expired.There was no external pressure on the femoral cannula.The customer did not exchange the cannula every six hours during the continuous ecmo due to contamination, miss-located position, or blood loss.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: report of damaged cannula device was confirmed.As received, the cannula was in two separated sections.A clean break was observed between the two separated sections of the cannula.The cannula was kinked and flattened along the wire-reinforced section of the cannula body.There was no other visual damage or other abnormality found.All vfem022 devices are 100% leak tested during manufacturing activities.It was reported that the cannula had been in use for 21 days which is beyond the short term use (6 hours or less) described by the ifu.There is no data to support use of this device beyond the 6 hours indicated on the ifu.A root cause cannot be determined at this time.The complaint trend was assessed and it was found to be in control.No further corrective or preventative actions are required at this time.Trends will continue to be monitored through the use of edwards quality systems and if action is required, appropriate investigation will be performed.
|
|
Search Alerts/Recalls
|
|
|