Catalog Number 2C4711K
Device Problem Material Perforation (2205)
Patient Problem No Patient Involvement (2645)
Event Date 03/22/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a tear in the filter of a small volume infusor was observed.The device was compounded with fluorouracil hs.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
It was reported that a tear in the filter of two (2) small volume infusors was observed.Manufacturing date -- november 17, 2016 ¿ november 18, 2016.Both devices were returned and evaluations are complete.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection was performed on both samples.One device had a tear in the filter and the other showed no evidence of a defect.The reported condition was verified for the first device and the cause is unknown.The reported condition was not verified for the second device.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
irvine CA
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6472073
MDR Text Key72111431
Report Number1416980-2017-03098
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4711K
Device Lot Number16M028
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/22/2017
Initial Date FDA Received04/07/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/12/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-