• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY PROFESSIONAL 30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY PROFESSIONAL 30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT; SCALER, ULTRASONIC Back to Search Results
Catalog Number 80395
Device Problems Leak/Splash (1354); Overheating of Device (1437)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received where this malfunction (overheating) resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a 30k fsi-sli-10s insert, the insert was leaking and hot after use; no injury resulted.
 
Manufacturer Narrative
The returned insert received from the practice was tested in the qa lab for the complaint of the insert is getting hot.The inserts was inspected and tested, the insert was tested for temperature and a maximum reading of 93.8 degrees was recorded.The insert was tested using a g-136 unit at 35cc of water flow, the test unit # was (b)(4), thermometer id # (b)(4) (cal date (b)(6) 2016 - cal due (b)(6) 2017).The insert has a severe leak at the distal from aggressive deflashing.In conclusion the complaint for the insert getting hot could not be duplicated and the insert could not be failed for the temperature due to needing a 118.4°f to warrant a failure.The insert can be deemed as failing for the leak issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key6472593
MDR Text Key72412599
Report Number2424472-2017-00031
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052334
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 05/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number80395
Device Lot Number16299
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/08/2017
Initial Date FDA Received04/07/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/25/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-