DENTSPLY PROFESSIONAL 30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT; SCALER, ULTRASONIC
|
Back to Search Results |
|
Catalog Number 80395 |
Device Problems
Leak/Splash (1354); Overheating of Device (1437)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
While no serious injury resulted in this event, there has been a previous report received where this malfunction (overheating) resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
While using a 30k fsi-sli-10s insert, the insert was leaking and hot after use; no injury resulted.
|
|
Manufacturer Narrative
|
The returned insert received from the practice was tested in the qa lab for the complaint of the insert is getting hot.The inserts was inspected and tested, the insert was tested for temperature and a maximum reading of 93.8 degrees was recorded.The insert was tested using a g-136 unit at 35cc of water flow, the test unit # was (b)(4), thermometer id # (b)(4) (cal date (b)(6) 2016 - cal due (b)(6) 2017).The insert has a severe leak at the distal from aggressive deflashing.In conclusion the complaint for the insert getting hot could not be duplicated and the insert could not be failed for the temperature due to needing a 118.4°f to warrant a failure.The insert can be deemed as failing for the leak issue.
|
|
Search Alerts/Recalls
|
|
|