• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERAKOS, INC CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS, INC CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Model Number NOT APPLICABLE
Device Problems Fluid/Blood Leak (1250); Hole In Material (1293)
Patient Problem Blood Loss (2597)
Event Date 03/16/2017
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction of the tubing leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review of kit lot f102 was conducted.There were no non-conformances related to the complaint.This lot met all release requirements.A review of kit lot f102 for the reported complaint issue shows no trends.Trends were reviewed for complaint category, tubing leak.No trends were detected for this complaint category.The assessment is based on information available at the time of the investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.(b)(4).Device not returned to manufacturer.
 
Event Description
The customer called to report a tubing leak that occurred during a treatment procedure.The customer stated that after 218ml of whole blood processed, a leak occurred due to a small hole in the tubing that surrounds the return pump.The customer reported that the treatment was aborted with no return of blood/products to the patient.The customer stated that the patient was in stable condition and did not require any medical intervention.The kit was not returned for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS, INC
hampton NJ
Manufacturer (Section G)
THERAKOS, INC.
10 north high street
suite 300
west chester PA 19380
Manufacturer Contact
megan vernak
po box 9001
53 frontage road
hampton, NJ 08827
MDR Report Key6474190
MDR Text Key72502965
Report Number2523595-2017-00062
Device Sequence Number1
Product Code LNR
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date01/01/2019
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXECP
Device Lot NumberF102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/16/2017
Initial Date FDA Received04/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/11/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age29 YR
Patient Weight61
-
-