Catalog Number 2C8750
Device Problem Fluid/Blood Leak (1250)
Patient Problem Blood Loss (2597)
Event Date 03/08/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
A clearlink blood recipient set leaked resulting in patient blood loss during an infusion.It was reported that the patient¿s body fluid was observed to be leaking onto the bed linen and the blood bag backed up.No further detail was provided regarding the type of infusion, the amount of blood loss, the patient¿s outcome or if medical intervention was required.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.During visual examination, a cut/hole was observed in the tubing.During functional examination, the leak was noted.The reported condition was verified; however, the cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina san cristobal
Manufacturer (Section G)
BAXTER HEALTHCARE - HAINA
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6476226
MDR Text Key72239131
Report Number1416980-2017-03137
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/04/2022
Device Catalogue Number2C8750
Device Lot NumberDR17A03037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/16/2017
Initial Date FDA Received04/10/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/11/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
-
-