Model Number M00545000
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
((b)(4).Reported event of wire broken.Investigation results: visual evaluation of the returned rx cytology brush found that the handle was slightly kinked.The device was in opened or extended position upon received.The bristled portion of the brush was present, properly formed and without issue.Functional analysis showed that the device failed to retract and the handle was disassembled finding that the pull wire had been kinked and broken at the hypotube joint.The reported complaint was confirmed, the pull wire broke.The review and analysis of all available information failed to indicate a root cause and is therefore, undeterminable.A review of the device history record (dhr) was performed and no deviations were found.A search of the complaint database revealed that no other complaints exist for the specified lot.
 
Event Description
It was reported to boston scientific corporation that an rx cytology brush was used during a procedure.According to the complainant, during the procedure, the brush broke inside the catheter.They also noticed that the device failed to extend and retract.There were no known patient complications reported as a result of this event.Attempts to obtain additional information regarding this event have been unsuccessful to date.Should additional relevant details become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX CYTOLOGY BRUSH
Type of Device
ENDOSCOPIC CYTOLOGY BRUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer (Section G)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6479514
MDR Text Key72442483
Report Number3005099803-2017-00982
Device Sequence Number1
Product Code FDX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K930348
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2018
Device Model NumberM00545000
Device Catalogue Number4500
Device Lot Number19791619
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/16/2017
Initial Date FDA Received04/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/05/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-