The device underwent an investigation and is now complete on the light fiber product 3269.Sbs06, lot 1905129 reported and returned to us.The device history lot review: the lot meet the specification outlined for release.We have not received any similar complaints on this lot.We have not received any information of same or similar like events that pertain to this lot or on any other lot of this same or similar manufactured product.Sample analysis: it was detected that the outside of the fiber was melted.This fiber component was produced at a different manufacturer.It is unknown how this occurred.It is possible that the gladding of the fiber was damaged during production, therefore heat was able to build up.Conclusion: the lot reported meets the specification outlined for release and is considered an isolated incident.Please note that should any further information be received, we will submit a supplemental report.
|