Model Number 10-80-00
Device Problem Physical Property Issue (3008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/28/2017
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative confirmed the reported issue and replaced the varilock clamp.Subsequent functional verification testing was completed without issue and the unit was returned to service.The investigation is on-going.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device has not been returned.
 
Event Description
Livanova (b)(4) received a report that varilock guide of the s5 roller pump broke during maintenance.There was no patient involvement.
 
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).The claimed variolock clamp was returned to livanova usa for further inspection.Evidence of spills have been identified.Moreover, a pin was identified to be missing and it allowed the arm to sit higher than intended when locked in place, and to come off when unlocked.A photo of the defective variolock clamp was sent to livanova (b)(4) for further investigation.The variolock clamp was of an old design pattern.As corrective action, a change order has already been implemented.A review of the dhr could not identify any deviations or non-conformities relevant to the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6490044
MDR Text Key72794820
Report Number9611109-2017-00301
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/28/2017
Initial Date FDA Received04/13/2017
Supplement Dates Manufacturer Received12/15/2017
Supplement Dates FDA Received01/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/10/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-