HILL-ROM BATESVILLE CLINITRON AT*HOME; BED, AIR FLUIDIZED Back to Search Results
Model Number P0800010016
Device Problem Insufficient Information (3190)
Patient Problem Pressure Sores (2326)
Event Date 03/22/2017
Event Type  Injury  
Manufacturer Narrative
The hill-rom technician informed the account that hill-rom no longer services this device.The patient was hospitalized for treatment when a third ulcer opened; the new ulcer is a stage 3.The patient's ulcers are on his buttocks and are being treated daily by a nurse.Development of pressure ulcers is multifactorial and cannot be only attributed to performance of the surface.Risk factors include protein-calorie malnutrition, microclimate (skin wetness caused by sweating or incontinence), diseases that reduce blood flow to the skin, such as arteriosclerosis, or diseases that reduce the sensation in the skin, such as paralysis or neuropathy.Position changes are key to pressure sore prevention and treatment.These changes need to be frequent, repositioning needs to avoid stress on the skin, and body positions need to minimize the risk of pressure on vulnerable areas.Per the hill-rom user manual, poor fluidization; if fluidization is sluggish or uneven, notify your hill-rom representative.Fluidization is affected by the following: room temperature, humidity, the amount of materials, such as fluid, cells or cellular debris, which has escaped from the blood vessels and has been deposited in tissues or on tissue surfaces, restricted air circulation from blankets on the bed.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2014.It is unknown if the facility performed any other preventative maintenance on this bed.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the patient developed a new wound.The bed was located at the patient's home.This report was filed in our complaint handling system as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
BED, AIR FLUIDIZED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
erin padgett
1069 state route 46 east
batesville, IN 47006
8129312791
MDR Report Key6493268
MDR Text Key72837179
Report Number1824206-2017-00173
Device Sequence Number1
Product Code INX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other Caregivers
Type of Report Initial
Report Date 03/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other Caregivers
Device Model NumberP0800010016
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/22/2017
Initial Date FDA Received04/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/13/2006
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-