• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; N/A Back to Search Results
Catalog Number A1059
Device Problems Material Protrusion/Extrusion (2979); Insufficient Information (3190)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
The distributor noticed the helicoil protruding outwards.It was unknown when it occurred.There was no patient involvement.
 
Manufacturer Narrative
Investigation completed 4/27/2017.Method: -device history - work order# 135849/ lot# 161/ serial# (b)(4): the device history record for this unit shows that this part was manufactured on 01/23/16.A total of 33 were produced for this work order number and lot number.No abnormalities related to reported incident found nor were there any variances, mrr¿s or reworks associated with this work order number and lot number.No service history is on file for this device.Conclusion: in summary, the device in question was not received for evaluation after several documented attempts.Therefore, the root cause to the end users experience could not be determined.This complaint will be reopened should we receive product.
 
Manufacturer Narrative
Investigation completed 07/25/2017.The device history record for this unit shows that this part was manufactured on 01/23/2016.No abnormalities relate to reported incident were found nor where there any variances, mrr¿s or reworks associated with this work order# and lot#.No service history is on file for this device.The customer removed the helicoil and returned the actual component (skull clamp helicoils coming off during pressure testing) that were removed from skull clamps repaired or in question.This customer sent in 21- gs127¿s (helicoils) that were removed from skull clamps repaired or in question.Due to how these helicoils were removed the heli coil threads were damaged thus we are unable to perform a dimensional inspection.We are also unable to trace these parts back to the device that it was removed from as these 21 pieces were received in one bulk bag with no supporting information root cause is undetermined at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
N/A
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH 45227
Manufacturer Contact
rowena bunuan
311 enterprise drive
plainsboro, NJ 08536
6099362393
MDR Report Key6507666
MDR Text Key73512559
Report Number3004608878-2017-00118
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberA1059
Device Lot Number135849/161
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/04/2017
Initial Date FDA Received04/20/2017
Supplement Dates Manufacturer ReceivedNot provided
07/25/2017
Supplement Dates FDA Received05/23/2017
08/02/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-