• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS INC MERIT CUSTOM KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS INC MERIT CUSTOM KIT Back to Search Results
Catalog Number K09-04816J
Device Problem Air Leak (1008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/23/2017
Event Type  malfunction  
Manufacturer Narrative
The device has been returned for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The account alleges that during a coronary angiography procedure, air was noticed leaking into the manifold from rotator connection.The air was successfully managed by the clinician throughout the procedure.No patient injury to report.
 
Manufacturer Narrative
The suspect device was returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A review of the complaint database was performed and no similar complaints were found for this lot number.A review of the device history record was performed and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERIT CUSTOM KIT
Type of Device
MERIT CUSTOM KIT
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS INC
1600 merit parkway
south jordan UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS INC
1600 merit parkway
south jordan UT 84095
Manufacturer Contact
casey hughes ms, cqe, csqp
1600 merit parkway
south jordan, UT 84095
8013164932
MDR Report Key6511059
MDR Text Key73359593
Report Number1721504-2017-00083
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 03/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date12/31/2019
Device Catalogue NumberK09-04816J
Device Lot NumberH1070553
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/28/2017
Initial Date FDA Received04/21/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-