• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, MARLBORUGH MA MYOSURE REACH HYSTEROSCOPIC TISSUE REMOVAL SYSTEM; UTERINE TISSUE REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, MARLBORUGH MA MYOSURE REACH HYSTEROSCOPIC TISSUE REMOVAL SYSTEM; UTERINE TISSUE REMOVAL SYSTEM Back to Search Results
Catalog Number 10-401FC
Device Problem Metal Shedding Debris (1804)
Patient Problem Foreign Body In Patient (2687)
Event Date 03/22/2017
Event Type  malfunction  
Manufacturer Narrative
Device history record (dhr) review was conducted for the reported identification number.The lot was released meeting all qa specifications.Currently unable to establish a relationship or impact to the reported observation.Device evaluation and event problem codes: the returned device was received, visually inspected, and tested.Visual examination of trd noted light scribe marks on the cutting window rim.This visual inspection indicative the blade damage was due to the user handling of the device.(b)(4).
 
Event Description
It was reported the physician performed a myosure procedure for uterine tissue removal on (b)(6) 2017 and the physician noted the "cutting blade shaved metal into the cavity." the physician removed the shaving from the patient.No patient injury.The procedure was completed and the patient was discharged home.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYOSURE REACH HYSTEROSCOPIC TISSUE REMOVAL SYSTEM
Type of Device
UTERINE TISSUE REMOVAL SYSTEM
Manufacturer (Section D)
HOLOGIC, MARLBORUGH MA
250 campus drive
marlborough MA 01752
Manufacturer Contact
sidra piracha
250 campus drive
marlborough, MA 01752
5082638884
MDR Report Key6511650
MDR Text Key73734197
Report Number1222780-2017-00094
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date06/30/2019
Device Catalogue Number10-401FC
Device Lot Number16F20RB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/22/2017
Initial Date FDA Received04/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/07/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MYOSURE CONTROL UNIT-SERIAL # UNK; MYOSURE HYSTEROSCOPE-SERIAL # UNK
Patient Outcome(s) Hospitalization;
-
-