TERUMO CARDIOVASCULAR SYSTEMS CORP. ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 816300 |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/27/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Per user, after they powered on the system, the cursor on the ccm did not move.Therefore, the clinical engineer had no choice but to force a shut-down.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the touch panel of the central control monitor (ccm) failed.As a result, they changed it with another ccm.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Evaluation is in progress, but not yet concluded.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) observed that the touch screen had failed when attempting to move the cursor.The cursor¿s position would not move from the right edge of the screen.The pst determined that the customer¿s touch screen was the cause of the problem.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|