• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 816300
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/27/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Per user, after they powered on the system, the cursor on the ccm did not move.Therefore, the clinical engineer had no choice but to force a shut-down.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the touch panel of the central control monitor (ccm) failed.As a result, they changed it with another ccm.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Evaluation is in progress, but not yet concluded.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) observed that the touch screen had failed when attempting to move the cursor.The cursor¿s position would not move from the right edge of the screen.The pst determined that the customer¿s touch screen was the cause of the problem.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6512033
MDR Text Key73567267
Report Number1828100-2017-00199
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K151349
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 06/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816300
Device Catalogue Number816300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/01/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/19/2017
Initial Date FDA Received04/21/2017
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received05/24/2017
06/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/18/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-