Model Number 37800
Device Problems Migration or Expulsion of Device (1395); Material Deformation (2976); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Twiddlers Syndrome (2114); Therapeutic Response, Decreased (2271)
Event Type  Injury  
Manufacturer Narrative
Other applicable components are: product id 4351-35 lot# serial# (b)(4) implanted: (b)(6) 2016: product type lead product id 4351-35 lot# serial# (b)(4) implanted: (b)(6) 2016.Product type lead.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A manufacturing representative (rep) reported on behalf of a healthcare provider (hcp) that the patient's symptoms had returned.The hcp did an x-ray and saw the device wires tangled.The patient turned the device insider their abdomen and cause the leads to migrate.New leads were implanted.No further complications were reported/anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6516006
MDR Text Key73499321
Report Number3004209178-2017-08906
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169360686
UDI-Public00643169360686
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2017
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/20/2017
Initial Date FDA Received04/24/2017
Supplement Dates Manufacturer Received04/20/2017
Supplement Dates FDA Received10/03/2017
Date Device Manufactured06/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-