• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO PIN COLLET (2.0-3.2MM); ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO PIN COLLET (2.0-3.2MM); ARTHROSCOPE Back to Search Results
Catalog Number 4100125000
Device Problem Metal Shedding Debris (1804)
Patient Problem No Patient Involvement (2645)
Event Date 12/02/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Thirteen devices were received for evaluation; 10 events were confirmed during testing.Five devices were found to have corrosion.Four devices were found to have shattered bearings.One device was found to be affected by metal shavings from a pin.The reported event was not confirmed for 3 devices; the devices were found to be within specifications for the reported event.The devices are not repairable and were not returned to the user facility.There were no remedial actions taken.The devices are not labeled for single-use.
 
Event Description
This report summarizes thirteen malfunction events, in which the device was shedding metal debris.Three reported events had no patient involvement; no patient impact.Ten reported events had patient involvement with no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIN COLLET (2.0-3.2MM)
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key6526415
MDR Text Key74148719
Report Number0001811755-2017-01144
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported13
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number4100125000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/02/2016
Initial Date FDA Received04/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-