Catalog Number 4100125000 |
Device Problem
Metal Shedding Debris (1804)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/02/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Thirteen devices were received for evaluation; 10 events were confirmed during testing.Five devices were found to have corrosion.Four devices were found to have shattered bearings.One device was found to be affected by metal shavings from a pin.The reported event was not confirmed for 3 devices; the devices were found to be within specifications for the reported event.The devices are not repairable and were not returned to the user facility.There were no remedial actions taken.The devices are not labeled for single-use.
|
|
Event Description
|
This report summarizes thirteen malfunction events, in which the device was shedding metal debris.Three reported events had no patient involvement; no patient impact.Ten reported events had patient involvement with no patient impact.
|
|