BOSTON SCIENTIFIC - CORK JETSTREAM® XC ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 112264-001
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/06/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by manufacturer: it is indicated that the device will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
 
Event Description
It was reported the blades stopped spinning.A 2.1mm jetstream® xc atherectomy catheter was being used during an atherectomy procedure in the left leg.At the end of the treatment, it seemed like the device got caught on something and it would no longer spin or rex back.The device was removed from the patient and the procedure was completed with a stent implanted.No complications were reported and patient was fine.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was further reported the lesion was 85% stenosed with minimal calcification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM® XC ATHERECTOMY CATHETER
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI 
Manufacturer (Section G)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI  
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6529316
MDR Text Key74025146
Report Number2134265-2017-04006
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130637
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Model Number112264-001
Device Catalogue NumberPV31300
Device Lot Number0020364659
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/06/2017
Initial Date FDA Received04/28/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age51 YR
-
-