• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS 8000 CARDIOPLEGIA MONITOR; COMPUTER, BLOOD-PRESSURE-SARNS 8000 CARDIOPLEGIA MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS 8000 CARDIOPLEGIA MONITOR; COMPUTER, BLOOD-PRESSURE-SARNS 8000 CARDIOPLEGIA MONITOR Back to Search Results
Model Number 16414
Device Problems No Device Output (1435); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/09/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation is in progress, but not yet concluded.The field service representative (fsr) verified the reported issue.The cardioplegia (cpg) monitor will display temperatures but not pressures.He replaced temperature/pressure printer circuit (pc) board.The cpg monitor is within manufacturer¿s specifications and is ready for clinical use.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the cardioplegia monitor would not display pressures.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
During laboratory evaluation, the product surveillance technician (pst) installed a lab use only (luo) cardioplegia monitor and found that the returned temperature/pressure board caused the monitor to not display pressures.Poor solder connections on the board were also found.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 CARDIOPLEGIA MONITOR
Type of Device
COMPUTER, BLOOD-PRESSURE-SARNS 8000 CARDIOPLEGIA MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6530080
MDR Text Key74163191
Report Number1828100-2017-00207
Device Sequence Number1
Product Code DSK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960916
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16414
Device Catalogue Number16414
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/12/2017
Initial Date FDA Received04/28/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/08/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/15/1995
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-