The device manufacturer date provided in the initial report was incorrect.The correct device manufacturer date is october 6, 2015.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).The cp5 was returned to livanova (b)(4) for further investigation.During evaluation, the reported issue could not be reproduced.The device was functionally tested, tested in a climate chamber, and underwent "electric static discharge" testing.No deviations or abnormalities were identified.For safety reasons the cp5 is planned to be scrapped.As the issue could not be reproduced, a root cause was not determined.
|