• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP SARNS 8000 UNIVERSAL ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP SARNS 8000 UNIVERSAL ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER Back to Search Results
Model Number 16402
Device Problem Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/11/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The field service representative (fsr) inspected the unit and confirmed the occlusion knob was difficult to adjust.The fsr regreased the occlusion knob.The unit operated to the manufacturer¿s specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the pump occlusion was jammed, not working, spinning and not moving in and out.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.Per the field service representative (fsr), a ¿sticky substance¿ was found on the sliders of the roller pumps.This appears to not have been degraded grease from the main shaft/knob area.Cleaning of this area of the roller pump resolved the stiff operation of the occlusion mechanism.He re-greased occlusion knob.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 UNIVERSAL ROLLER PUMP
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6540521
MDR Text Key74484069
Report Number1828100-2017-00215
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K953901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16402
Device Catalogue Number16402
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/11/2017
Initial Date FDA Received05/03/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/04/1999
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-