Model Number EX1150
Device Problems Thermal Decomposition of Device (1071); Failure To Service (1563)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/31/2017
Event Type  malfunction  
Event Description
It was reported the physician was having trouble working with the merlin transmitter as there was no power being generated for a transmission.The transmission still could not be retrieved when using a second power outlet.The physician decided to remove the prongs and the patient noted the prongs were discolored and charred on the left side.The physician replaced the transmitter.
 
Manufacturer Narrative
The reported field events of no transmitter power and burn marks on the prongs were confirmed in the laboratory.Visual inspection noted burnt components on the main circuit board of the ac power adapter as well as the inner housing right behind the main pcb of the transmitter.It is believed that the root cause of the burnt components was high current flow through the telephone line and through the ac wall power outlet which lead to the transmitter not being able to power on.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6544373
MDR Text Key74419664
Report Number2938836-2017-23417
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 08/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Other Device ID Number05414734504799
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/31/2017
Initial Date FDA Received05/04/2017
Supplement Dates Manufacturer ReceivedNot provided
08/01/2017
Supplement Dates FDA Received06/12/2017
08/09/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-