MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 8DCT
Device Problem Air Leak (1008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/14/2017
Event Type  malfunction  
Manufacturer Narrative
Are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received a report that the endotracheal tube had an air leak.The site stated they were able to change the endotracheal tube with no ill effect to the patient.The site stated there was no patient injury.
 
Manufacturer Narrative
The sample was received for analysis and the reported issue was confirmed.A inflation/deflation test was performed and it was observed the cuff deflated immediately.A visual inspection was performed and it was observed the cuff presented two small tears.Manufacturing controls are in place to detect cuffs with damage and to reduce the potential for occurrence during the manufacturing process.Information has been added to the database and trends will continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Type of Device
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX   32590
Manufacturer Contact
sharon murphy
15 hampshire street
mansfield, MA 02048
2034925267
MDR Report Key6546382
MDR Text Key74477032
Report Number2936999-2017-05187
Device Sequence Number1
Product Code JOH
UDI-Device Identifier20884522029022
UDI-Public20884522029022
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962173
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2021
Device Model Number8DCT
Device Catalogue Number8DCT
Device Lot Number16C0796JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/14/2017
Initial Date FDA Received05/04/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-