The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.Microscopic analysis of the returned delivery system revealed clearly visible stent embedding traces on the balloon surface, indicating that the stent was properly crimped between the two x-ray markers at the time of delivery.Review of the manufacturing history of the product detailed above did not reveal any non-conformity.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.The device in question was manufactured according to specifications and successfully passed all in-process controls as well as the final inspection.Based on the conducted investigations no material or manufacturing related root cause could be determined.The final root cause for the stent dislodgement is most likely related to external factors during intervention.Further, the stent was used in the posterior tibial artery against the ifu.
|