TERUMO CARDIOVASCULAR SYSTEMS CORP. ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 6382 |
Device Problem
Bent (1059)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/13/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Evaluation is in progress, but not yet concluded.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, there was flow sensor with a bent pin.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed a bent pin in the connector (pin 6) of the flow sensor.Per pst after proving the proper operation of a centrifugal system water loop and flowmeter circuit using lab-use components, the device under test (dut) flow sensor provided no reading.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|