• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problems Low Readings (2460); Device Operates Differently Than Expected (2913)
Patient Problems No Consequences Or Impact To Patient (2199); Low Cardiac Output (2501)
Event Date 04/27/2017
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the companion 2 driver continued to perform its life-sustaining functions.The companion 2 driver will be returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the companion 2 driver exhibited diagnostic waveform delay with decreased cardiac output while supporting a patient.The customer also reported that the patient was switched to a backup driver without any adverse patient impact.
 
Manufacturer Narrative
The companion 2 driver was returned to syncardia for evaluation.The customer-reported issues were reproduced during investigation testing; the driver exhibited delayed diastolic waveform timing and decreased cardiac output when connected to an external air source.The root cause was determined to be a malfunction of the left and right pilot valves.Syncardia has an open corrective and preventive action (capa) for this issue.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Event Description
The customer, a syncardia certified hospital, reported that the companion 2 driver exhibited diagnostic waveform delay with decreased cardiac output while supporting a patient.The customer also reported that the patient was switched to a backup driver without any adverse patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key6552713
MDR Text Key74794932
Report Number3003761017-2017-00073
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 04/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/27/2017
Initial Date FDA Received05/08/2017
Supplement Dates Manufacturer Received08/09/2017
Supplement Dates FDA Received08/17/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age59 YR
-
-