• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH SIREMOBIL COMPACT; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH SIREMOBIL COMPACT; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number 3776494
Device Problem Insufficient Information (3190)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/10/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.(b)(6).
 
Event Description
It was reported to siemens that a malfunction occurred following use of the siremobil compact system.The customer reported the image and monitor are black and it is not possible to see objects placed on the image intensifier.There is no report of impact to the state of health of any patient involved.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a defective power supply hv30-2 (pn:03830580).With the defective power supply, the image intensifier is powerless and cannot work.As a result, the image remains black, as mentioned in the complaint.The spare parts consumption of the power supply has been checked and no error accumulation has been recognized.The power supply has been exchanged on site by the local service organization and the error did not reoccur.The manufacturer is not considering further actions resulting from this event as no general systematic error has been recognized.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIREMOBIL COMPACT
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemensstrasse 1
forchheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forcheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
65-1a
malvern, PA 19355
6104486461
MDR Report Key6556483
MDR Text Key75003893
Report Number3004977335-2017-77304
Device Sequence Number1
Product Code JAA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K040066
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 04/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number3776494
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location Hospital
Date Report to Manufacturer04/12/2017
Initial Date Manufacturer Received 04/12/2017
Initial Date FDA Received05/10/2017
Supplement Dates Manufacturer Received04/10/2018
Supplement Dates FDA Received04/10/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-