• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM Back to Search Results
Model Number 6379
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/18/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation is in progress, but not yet concluded.The field service representative (fsr) confirmed the display randomly turned off, blinking and resetting the unit.The membrane panel was separating causing a short that reset the unit.The pump display / membrane panel assembly was replaced.The unit operated to the manufacturer¿s specifications.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the centrifugal unit turns off randomly.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) was not able to duplicate the reported complaint.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS CENTRIFUGAL SYSTEM
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6563912
MDR Text Key75160457
Report Number1828100-2017-00232
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6379
Device Catalogue Number6379
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/18/2017
Initial Date FDA Received05/12/2017
Supplement Dates Manufacturer Received08/24/2017
Supplement Dates FDA Received09/06/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-