Model Number 840 |
Device Problem
Occlusion Within Device (1423)
|
Patient Problem
No Code Available (3191)
|
Event Type
Injury
|
Manufacturer Narrative
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that while in use on a patient, a 840 ventilator had a severe occlusion.There was no harm to the patient.A service engineer (se) inspected the device and found that the patient circuit was missing as well as that the exhalation filter was cracked.The exhalation valve was found to be noisy during flow sensor calibration.To address the failure, the se replaced the exhalation filter and the exhalation check valve.The unit passed all testing and operates within the manufacturing specifications.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|