Catalog Number 701048012
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).According to service order (b)(4) the venous probe teaching procedure was performed using the cardiohelp service tool software.It is now working properly and initializes as it should.This unit passed all functional and safety tests.It has been returned to the customer and cleared for clinical use.(b)(4).A supplemental medwatch will be submitted if new information has been received.
 
Event Description
(b)(4).The venous probe failed to initialize during patient transport.The cardiohelp was switched out.It was reported that there was no adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOHELP SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key6572308
MDR Text Key75430260
Report Number8010762-2017-00157
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K102726
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number701048012
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/02/2017
Initial Date FDA Received05/17/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-