TERUMO CARDIOVASCULAR SYSTEM CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 802100
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/26/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation is in progress, but not yet concluded.The field service representative (fsr) was able to get the touch screen to work but it was intermittent.The fsr replaced the unit with another ccm and the suspect unit was returned to the manufacturer for further evaluation.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the touch screen of the central control monitor (ccm) was not working.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The product surveillance technician (pst) conducted an evaluation of the returned device, and the complaint was duplicated.The touch screen was not working at gouged areas.Exterior inspection revealed multiple gouges on touch screen.He connected the ccm to aps1 power supply and powered the device on.The pst attempted to move the cursor to different areas of touch screen.It was determined that the touch screen was the cause of the problem.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEM CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6577952
MDR Text Key75624552
Report Number1828100-2017-00248
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151349
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number802100
Device Catalogue Number802100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/26/2017
Initial Date FDA Received05/19/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/09/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse