Model Number 802100 |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/26/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Evaluation is in progress, but not yet concluded.The field service representative (fsr) was able to get the touch screen to work but it was intermittent.The fsr replaced the unit with another ccm and the suspect unit was returned to the manufacturer for further evaluation.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the touch screen of the central control monitor (ccm) was not working.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The product surveillance technician (pst) conducted an evaluation of the returned device, and the complaint was duplicated.The touch screen was not working at gouged areas.Exterior inspection revealed multiple gouges on touch screen.He connected the ccm to aps1 power supply and powered the device on.The pst attempted to move the cursor to different areas of touch screen.It was determined that the touch screen was the cause of the problem.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|