BIOMET ORTHOPEDICS VERSA-DIAL COMPREHENSIVE STANDARD ADAPTOR; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problems Migration or Expulsion of Device (1395); Unstable (1667)
Patient Problems Pain (1994); No Code Available (3191)
Event Date 12/21/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.The following sections could not be completed with the limited information provided.Date explanted ¿ n/a.Concomitant devices ¿ versa-dial modular head with variable offset catalog #: 113055 lot #: 705460, hybrid; glenoid base 4mm catalog #: 113954 lot #: 985500, pt hybrid glenoid post catalog #: pt-113950 lot #: 762830.This report is number 1 of 4 mdrs filed for the same patient (reference 0001825034-2017-03361 / 03362 / 03363).
 
Event Description
It is reported that the patient underwent a total shoulder arthroplasty.Subsequently, pain, tenderness, impingement, subacromial crepitus, and a infraspinatus fosia deformity were noted at six (6) week post-operative follow-up.Continued pain, biceps tendon tenderness, impingement, infraspinatus atrophy, and instability were noted at three (3) month post-operative follow-up.Pain, tenderness, impingement, and instability were noted at one (1) year post-operative follow-up.Tenderness and instability were still noted to persist at two (2) year post-operative follow-up.Lastly, a small lucency at the inferior aspect of the implant was noted at thirty-four months post-operatively.However, the condition is noted to be tolerated and there is no plan to revise the patient or otherwise treat them.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of x-rays provided.X-ray reviewer stated "imaging findings are thought to confirm the reported event.Patient risk factors (acromioclavicular hypertrophy and cortical irregularity anterior humeral head), as well as imaging findings of anterior / superior subluxation of humeral head known are known to be associated with reported events of supraspinatus / greater tuberosity tenderness, biceps tendon tenderness (or rupture), impingement and instability." device history record (dhr) was reviewed and no discrepancies were found.Review of the complaint history determined that no further action(s) is/are required.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSA-DIAL COMPREHENSIVE STANDARD ADAPTOR
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6580345
MDR Text Key75599481
Report Number0001825034-2017-03360
Device Sequence Number1
Product Code MBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK060716
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date11/30/2023
Device Model NumberN/A
Device Catalogue Number118001
Device Lot Number303560
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/21/2017
Initial Date FDA Received05/20/2017
Supplement Dates Manufacturer Received07/28/2017
Supplement Dates FDA Received08/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/20/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age71 YR
Patient Weight81
-
-