The balloon burst was confirmed.Two comparative catheters were tested for rated burst pressure.The labeled rbp for this catheter is 4.0 atm.The comparative catheters were taken to 8.0 atm and 7.0 atm, respectively, before burst.The user facility stated that they did not use an inflation device with pressure gauge to inflate this balloon as is recommended in the instructions for use.Due to them not using an inflation device with pressure gauge, it is unknown as to what pressure they took the balloon to.It is likely the balloon burst due to over inflation.The balloon was inflated with 17.5ml of dye.Hospital stated this volume was according to company booklet (instructions for use).However, numed does not recommend nor provide a volume chart for this product.The instructions for use states to use an inflation device with pressure gauge and lists rated burst pressures for this catheter.The patient is doing fine.
|
As reported by baylis medical: " dr.(b)(6) was using the nucleus balloon for aortic valvuloplasty for pre-tavi procedure, they uses 17.5 ml of dye to inflate the balloon.During the inflation of the aortic valve, the balloon ruptured and he los 2/3 of the balloon into the patient which they never found.Balloon ruptured and they lost pieces in the body that they didn't find.".
|