Model Number SC-2158-70 |
Device Problems
Bent (1059); Fracture (1260); High impedance (1291); Kinked (1339); Migration or Expulsion of Device (1395)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 04/12/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: model#: sc-2158-70 serial#: (b)(4) description: linear lead with enhanced stylet, 70cm model#: sc-4318 lot#: 19872195 description: clik x anchor.
|
|
Event Description
|
A report was received that the patient was experiencing lost of coverage.It was also reported that the leads had migrated and a lead was having four contacts out.The patient underwent a revision procedure wherein two leads and click anchors were replaced.Device malfunction was suspected.The patient was reportedly doing well postoperatively.
|
|
Manufacturer Narrative
|
Sc-2158-70 sn: (b)(4) device evaluation indicated that the lead passed all tests performed.Sc-2158-70 sn:(b)(4) device evaluation indicated that the complaint has been confirmed.Visual (microscope) and x-ray inspection of the lead revealed that multiple cables were completely broken at the bent/kinked location of the lead.The bent/kinked location is 3 cm from the set screw mark of the clik x anchor.There are no exposed cables at the clik site fracture locations.The broken cables resulted in the reported complaint of high impedance.
|
|
Event Description
|
A report was received that the patient was experiencing lost of coverage.It was also reported that the leads had migrated and a lead was having four contacts out.The patient underwent a revision procedure wherein two leads and click anchors were replaced.Device malfunction was suspected.The patient was reportedly doing well postoperatively.
|
|
Search Alerts/Recalls
|